Equipment Validation Protocol for the Pharma Manufacturing Industry – What’s contained in it? 

By: Gerry Creaner B.Chem Eng and Donagh Fitzgerald B.Prod Eng. Last Updated: July 2024

https://www.getreskilled.com/wp-content/uploads/2020/05/What-is-an-IQ-OQ-an-PQ-validation-protocol-getreskilled.com_.jpghttps://www.getreskilled.com/wp-content/uploads/2020/05/What-is-an-IQ-OQ-an-PQ-validation-protocol-getreskilled.com_.jpg

An Equipment Validation Protocol (also known as an equipment qualification protocol) is a written plan that outlines how to test and verify equipment, piping, instruments and utilities in a critical system to make sure it works properly and consistently according to its intended use.

The protocol is used in regulated industries such as pharmaceutical manufacturing and typically includes these processes:

  1. Installation Qualification (IQ): Verifies that the equipment is installed correctly, according to manufacturer’s specifications and site requirements.
  2. Operational Qualification (OQ): Tests the equipment to ensure it operates as intended within established parameters.
  3. Performance Qualification (PQ): Confirms that the equipment performs under load according to the design specifications.

Each step (IQ, OQ, PQ) builds on the previous one, progressively confirming the equipment’s reliability and functionality. These steps are collated in the equipment validation protocol which:

  • lists all the equipment, piping, instruments and utilities that need to be tested
  • defines the test scripts that must be followed to make sure the equipment operates correctly
  • outlines how to perform the tests
  • specifies who will conduct them
  • records whether the equipment, along with its piping and instruments, passes or fails the tests
  • details the required proof and the expected outcomes
  • describes the steps to take if the actual outcomes deviate from expectations

Each critical system has its own protocol.

The objective is to create documented evidence to prove “fitness for use ” of an equipment system. It makes sure that a facility and the equipment in it, function as required so that it can be approved by the regulatory agencies that have jurisdiction over that facility.

Note: Equipment Validation and Equipment Qualification are often used interchangeably. However, some industry professionals argue that ‘Equipment Qualification’ is more precise, as you ‘Qualify’ equipment, piping, instruments, and utilities. In contrast, ‘Validation’ is typically reserved for processes, procedures, and methods. For the purposes of this guide, we will use both terms.”

Equipment Validation Training Course

If you need to learn how to populate an IQ IQ PQ equipment qualification protocol, check out our Equipment Validation (IQ OQ PQ) Training Course – For Starter Validation, CQV and C&Q Roles.

Content

We will do a deep dive into what’s contained within a validation protocol template and guide you through the process of writing one. We’ll also take the example of validating a clean-in-place system (CIP) within a pharmaceutical plant using the relevant supporting documents to demonstrate how to complete the template.

But before we start, I’d strongly recommend you carefully read the following two articles to give yourself an overview of the validation and qualification process and familiarise yourself with some of the relevant terminology.

Determine the Qualification Requirements

Most validation projects will take an existing blank template and modify it for the project at hand rather than create a new blank template from scratch every time. But there may be times where you’d need to create key pieces of the protocol from scratch.

The protocol will detail the following elements.

  • Product characteristics – showing what your system is looking to achieve/produce
  • Production equipment – detailing the equipment necessary
  • Test scripts and methods – telling you the steps involved in conducting a test
  • Test parameters and acceptance criteria – defining acceptable test results
  • Test checksheets – documenting and recording the test results
  • Final approval – documenting that the validation process has been successfully carried out

What are the Key Elements of a Qualification Protocol?

Think of this as preparing a highly detailed checklist that details factors like:

  • Impact assessments (risk assessments) – you don’t want to validate and test everything as that would be hugely expensive and wasteful. So first you conduct a component-level impact assessment on the system to figure out what components are critical and directly impact product quality vs those items which are non-critical and don’t and only validate the items that are critical.
  • IQ OQ PQ test checksheets – e.g. equipment installation, instrumentation installation, piping installation, P&ID walkdown installation checksheets to document and record results
  • Test scripts and methods – telling you the steps involved in conducting a test
  • Test parameters and acceptance criteria – defining acceptable test results
  • Product characteristics – showing what your system is looking to achieve/produce
  • Production equipment – detailing the equipment necessary
  • Final approval – documenting that the validation process has been successfully carried out

For this example, we are going to describe the key elements contained within a blank protocol for a Clean in Place (CIP) System along with the key thinking and intent behind each section. The blank template is about 60 pages in length. By the time it has been filled out, it could stretch to 150 pages or more, depending on the size of the project.

It’s worth keeping in mind that the protocol is laid out for the convenience of those who follow, not for those who are initially filling it in. Sections are laid out to walk someone through the completed validated system, not in the order you’d fill them in.

Note: While we can’t share the full blank protocol with you (as it’s proprietary material), we will share some key elements of the protocol in the form of blank templates that you can download for your own use.

1 – Approvals Page

Approvals Page GetReskilled

On the protocol approvals page, you would complete a list of necessary protocol approvers – the people who need to “sign off” on the validation activities.

How this would “look” in the real world:

This section is where validation team members have their responsibilities summarised, and sign to state that everything they’ve written in the document is correct and accurate. Each of the relevant personnel would be listed on this page.

While this is the first page of a protocol, it is the last page that will be signed. Signatures on this page confirm that everything in the rest of the document is accurate and that you are happy with the results and recommendations.

You should never sign anything unless you know it to be an accurate reflection of the situation. You would never sign anything that was not your own work. Always remember that in an FDA audit, the auditor could ask to speak to you about your specific role in the validation process, and you would have to be able to defend the decisions you made.

Additional Reading

What Types of Jobs are there in the Pharmaceutical Industry?
What Is a Validation Engineer?

2 – System Description

System Description GetReskilled

In this section, you need to prepare a summary ‘System Description’. You might include things such as:

  • Drawing a process flow / block diagram
  • Listing critical operational features
  • Describing the major equipment components in the context of their installation and functional features.

How this would “look” in the real world:

In an FDA audit, this is where the auditors will go first to understand the system being validated. It provides a high-level overview of the equipment system. It needs to be short, but also provide the level of detail they need to understand the rest of the protocol.

When you first begin in a Validation team, this will probably already be completed for you. However, as you gain seniority, you will need to understand the purpose of a system description and be able to write one of your own.

3 – Scope

In our example, we have broken this into 3 parts.

System Impact Assessment:

System Impact Assessment GetReskilled

In this section you describe why the CIP System is considered a ‘direct impact’ system.

Component Impact Assessment:

Summary Table Operational Critical Components GetReskilled

For the identified equipment, instruments, and piping, you must determine if the components are critical or non-critical – and if critical, whether they are product-contact critical, operationally critical, or both.

Let’s take a look at how you differentiate between by product-contact critical items, operationally critical items, and non-critical items

Product-contact critical items come into direct contact with the product OR are part of the chain of surfaces through which there is a risk of contamination being carried to the product.

The first part of that is quite self-explanatory but the second part can take a bit of thinking about. So let’s consider an everyday example…

You’re in the kitchen and you cut raw chicken. Once you’ve finished, you don’t wash your hands. Instead you go to the fridge and take out salad items. You cut them, and serve the salad to your friend. Your friend gets ill from salmonella after eating the salad.

Your friend was never near the raw chicken but there was a chain of surfaces (e.g. the skin on your hands, and the outer surface of the salad items) that brought the salmonella directly from the raw chicken to your friend.

Our clean in place system has that type of product-contact critical items. The final medicine or its ingredients never come near the clean in place system but the system is part of a chain of surfaces that can pass contamination to the product.

Think of the detergent tank of our system. The inner surface of the tank touches the detergent liquid. That detergent liquid passes along a system of pipes and into the reactors during a drawdown. That same liquid then touches the inner surface of the reactor. After a cleaning cycle is finished and manufacturing restarts, the inside of that reactor will come into direct contact with product ingredients.

So although no final products (or even ingredients) are ever in direct contact with the inside of the detergent tank, there is a chain of surfaces that leads from the tank to a surface that will come into contact with the product. If there was contamination in the detergent tank, there is a possibility it could ultimately reach the product via this chain of surfaces. Therefore, our detergent tank is product-contact critical.

Operationally critical items don’t come into direct contact with the product and they’re not part of that chain of surfaces that can introduce contamination. But they contribute to maintaining a state of balance (dynamic equilibrium) for the key factors (e.g. temperature, flow, level, concentration) in our Clean in Place system.

If any of these 4 key factors are out of specification in our CIP system (e.g. liquid is not hot enough or detergent concentration isn’t strong enough), we cannot guarantee that the system is working as intended. Operationally critical items are involved in controlling, monitoring, or reporting these 4 factors so it’s essential that we validate them.

(Note that these key factors may be different depending on the system you’re validating, the 4 listed are specific to controlling our CIP system.)

Non-critical items do not come into contact with the product, nor can they influence the factors that are essential for the correct functioning of the system.

Summary Table – Operational Critical Components:

Component Impact Assessment GetReskilled

For any component identified as operationally critical, you’ll then complete the summary table of the operational critical components and include rationale.

How this would “look” in the real world:

When you begin in the field – the component impact assessment will probably have been completed for you . However, over time, you are expected to be able to conduct an impact assessment on your own. This is one of the most critical parts of a protocol.

The examples above and your knowledge of ISPE Guide Baseline 5 will be key when it comes to making decisions about criticality, and what needs validation as a result.

It’s important to remember the remit of the FDA at this point and instil it in your work ethic – to ensure the “manufacture safe medicines” and “at an affordable cost”.

You should always keep in mind that the more items that need to be validated, the costlier the validation process will be, and the costlier the medicine will be for the patient. You need to validate the key items, but you cannot just validate everything “just incase”. It’s this balancing act you will need to perform in the field.

You have to be able to justify your decisions around which things you deemed critical if questioned by an FDA auditor.

Always remember, that while an FDA auditor might hold a different opinion on which items are critical, they will respect your decisions if you use scientific-based risk management tools to reach them. There can be different opinions as there is often no single right answer to what is critical and what is not. There are wrong answers of course, but many decisions on critical items are shades-of-grey rather than black or white.

4 – P&ID Scope

P&ID Scope

In this section you include a copy/scan of the P&ID and either draw a ‘scope bubble’ or yellow-highlight the ‘in-scope’ components.

How this would “look” in the real world:

This section builds on the decisions you made earlier. In the field, the decisions you made regarding the scope of the protocol (i.e. what equipment elements, instrumentation components, and line and piping elements are part of the system you are validating) determine what you include in the scope bubble here. Again, as a new member of a validation team, you will likely be told what is included but over time you will be expected to make the decisions yourself.

5 – Definitions / Glossary

Definitions: Glossary GetReskilled

Make a list of abbreviations that are used in the protocol and their definitions. This is something you should fill up as you complete the protocol. Once you have finished, read back through the protocol and make sure to add any abbreviations you have missed.

How this would “look” in the real world:

The purpose of this section is so that the person who comes along after you can understand any abbreviations you have used throughout the protocol. We all routinely use some abbreviations, this is no different in companies. But an abbreviation used on the factory floor that people around you understand is not necessarily something an outsider or someone new to the company will know. When it comes to an FDA auditor, they need to be able to understand exactly what you are referring to with your abbreviations – so make sure this is as comprehensive as possible.

6 – Responsibilities / Subject Matter Experts

Responsibilities / Subject-Matter Experts GetReskilled

In this section, you identify subject matter experts. These are the people responsible for the development, review, approval, and execution of the protocol. Complete a list of project responsibilities for each of the specific subject matter experts.

How this would “look” in the real world:

When you begin in a Validation Team, you will be told exactly who needs to sign off on your work and what responsibilities each team member has. If you go on to be a Validation Engineer and begin to write protocols from scratch, you could then be responsible for deciding what each department in the company is responsible for, who the subject matter experts are, and what responsibilities each person has.

An FDA auditor will turn to this protocol section to see each department or individual’s responsibilities so they know who to talk to if they have specific questions about a test or procedure. It is therefore important that this section is detailed and accurate.

This is also one of the sections that is closely linked to other protocol sections. Your Approvals Page (Section 2), Your Responsibilities Page (Section 10) and Your Progression Approval Page (Section 12), should all align. If they do not, an auditor is going to have serious questions.

7 – Calibration Program Verification

Responsibilities / Subject-Matter Experts GetReskilled

In this section, you need to give a summary of the ‘Calibration Program Verification’ procedure that is used to guide the creation of calibration check sheets (these are usually located in the protocol appendix). Make sure you write this from the perspective of someone who is entirely new to the system you are validating.

How this would “look” in the real world:

After you have identified the Operationally Critical Components (remember we did this in Section 3), you need to make sure that any operationally critical instruments you have identified become part of the maintenance team’s calibration rota.

Calibration is where you test to make sure the operationally critical instruments are working. For example, you would test to make sure that the temperature gauge is measuring the temperature correctly. This is done on a routine and ongoing basis to make sure everything is operating as intended for the whole length of the system life cycle. Commissioning, Qualification & Validation (CQV), Commissioning & Qualification (C&Q) Engineers and Maintenance personnel are responsible for calibration testing in a plant. It falls into “Continued Process Verification”, the purpose of which is to ensure that the previously validated system remains in a reliable and validated state.

Your job in the field would be to write this summary, then use the information from this to create a checksheet to be completed during calibration activities. Finally, the CQV Engineers and Maintenance Personnel would be told to add this list of instruments to their regular calibration schedule.

Additional Reading

8 – Preventative Maintenance Program Verification

Preventative-Maintenance Program-Verification-GetReskilled

In this section, you need to give a summary of a ‘Preventative Maintenance Program Verification’ procedure that is used to guide the creation of a check sheet (that is usually located in the protocol appendix). Make sure you write this from the perspective of someone who is entirely new to the system you are validating.

How this would “look” in the real world:

Each plant has a preventative maintenance program. Preventative maintenance is how you ensure that all the operationally critical components you have identified, are working properly at all times.

The preventative maintenance team performs maintenance checks on equipment before there is a problem – like changing the tires on your car before they become bald. On a car, your tires should be good for 25,000 miles, and therefore if you change them at 20,000 miles they should have enough thread to be safe to drive on. The preventative maintenance system identifies the critical items that need to be maintained, and the duration between each maintenance check.

CQV Engineers, C&Q Engineers and Maintenance Personnel are responsible for the preventative maintenance program and it is your job here to complete the form with what components they need to add to preventative maintenance rota

Preventative Maintenance is also part of the stage of Validation called Continued Process Verification. Failure to complete this stage of validation could result in compromised batches of unreliable quality medicine that are not safe for patients to use.

The activities of this stage focus on two key goals:

  1. To maintain equipment and facilities in their validated state
  2. To quickly highlight variation in product or process data, allowing corrective action to be taken

Activities involved in Continued Process Verification might include things such as:

  • Regular equipment maintenance
  • Routine testing of in-process materials and final product sampling (comparing both within and between batches)
  • Routine gathering and analysis of in-process data
  • Equipment swabbing
  • Cleaning

Additional Reading

Process Validation

9 – Draft SOP (Standard Operating Procedure) Verification

9 Draft-SOP-(Standard-Operating-Procedure)-Verification-GetReskilled

In this section, you’ll propose a list of necessary standard operating procedures for the ‘Clean in Place’ system.

The information you outline here will be used to create an SOP checksheet in the protocol appendix.

How this would “look” in the real world:

As you know by now, no matter what role you have in the Pharma or MedTech industries, a key part of your job will always involve following strict internal protocols and Standard Operating Procedures (SOPs) as well as following current Good Manufacturing Practices (cGMP).

When you start in a Validation Team, you are going to be told which SOPs are going to be implemented because of the validation protocol that has been written. As you go up the career ladder, you will be expected to make the decision regarding what SOPs are needed.

This section basically tells an auditor that at this stage of the protocol, you were thinking of implementing the following list of SOPs to make sure the equipment functions as intended. When you get to later sections, you are then going to tell an auditor (and your colleagues) the finalised list of SOPs that are to be followed.

Writing SOPs is an advanced-level task but it’s important that you understand where they come from and how they integrate with the rest of the protocol.

Additional Reading

How to Save Money with SOPs that Stand the Test of Time

10 – Test-Instrument Calibration Verification

Test-Instrument-Calibration-Verification-GetReskilled

In this section, suggest a list of test instruments that may be required for the testing phase.

The information you outline here will be used to create a test-instrument calibration verification checksheet in the protocol appendix.

How this would “look” in the real world:

This section is very simple. You want to document that all test instruments used to validate the readings on the instruments of the Clean in Place system are functioning correctly. We’ve already talked about the importance of calibration but in this case, in the future, if there was a problem with any of the instrument readings on the CIP system, you want to rule out the possibility that it was a problem that stemmed from a faulty test instrument reading

11 – Progression Approval to Proceed to Functional Verification

Progression Approval to Proceed to Functional Verification GetReskilled

Align the signatories in this section “Progression Approval to Proceed” with the section 2 signatures on the ‘Approvals Page’, section 6 Responsibilities / Subject Matter Experts, and the protocol signature log which will be contained in the appendix.

How this would “look” in the real world:

This section is where you would sign your name stating that everything you have written is correct and accurate as you move into functional verification. You’ll see again that all relevant personnel have to sign at this stage, and responsibilities for each are summarised.

Remember, you should never sign anything unless you are sure it’s an accurate reflection of the situation – and you would never sign anything that was not your own work. In an FDA audit, where an auditor was looking at this specific piece of equipment, the auditor could ask to speak to you about your specific part validating the equipment, and you could have to justify decisions you made, to them.

This is a section that builds on what you have written previously. What is written here should match what you wrote in section 2 and 10. And again, an auditor would have questions for you if it didn’t.

12 – Operational Critical Instrument Calibration Verification Form

Operational Critical Instrument Calibration Verification Form GetReskilled

Enter the same list of instruments into the ‘Operational-Critical-Instrument Calibration Verification Form’ in this section as those listed in Section 3 ‘Summary Table – Operational Critical Components’.

How this would “look” in the real world:

The purpose of this form is to ensure that critical instruments associated with the system are in a calibrated state during the operational qualification (OQ) phase.

Operational Qualification is used to confirm that the overall CIP system will operate as intended

To carry this out, C&Q engineers must carry out various tests on the equipment to make sure the system functions is meeting the expectations outlined in the User Requirement Specification (URS) for flow, temperature etc.

13 – Functional Verification Testing Procedure

Prepare functional test procedures for Temperature, Flow and Level for the Clean in Place System.

13 Functional-Verification-Testing-Procedure-GetReskilled

The sections included in your test procedures should be:

a) Short Description of the test
b) Linkage to the requirements challenged (and operationally critical components)
c) Acceptance criteria for the test
d) Prerequisites and or assumptions
e) Specific Test steps
f) Expected results and actual results

Use those to prepare complimentary functional verification check sheets for Temperature, Flow and Level – as critical features of the Clean in Place System.

13 Functional-Verification-Testing-Procedure-GetReskilled

These check sheets will be located in the Appendix.

How this would “look” in the real world:

When you first begin in a Validation Team, you will be given the list of tests you have to perform, the steps involved in doing so, and what the acceptable range of results are. However, as you move up the career ladder, you will be expected to write these OQ tests and OQ GMP checksheets by yourself.

14 – Approved SOP Verification

Approved SOP Verification GetReskilled

In this section you list out the approved SOPs.

How this would “look” in the real world:

As you know by now, no matter what role you have in the Pharma or MedTech industries, a key part of your job will always involve following strict internal protocols and Standard Operating Procedures (SOPs) as well as following current Good Manufacturing Practices (cGMP).

When you start in a Validation Team, you are going to be told which SOPs are going to be implemented as a result of the validation protocol that has been written. As you move up the career ladder, you will be expected to make decisions regarding which SOPs you think are needed.

This section tells an auditor (and your colleagues) the finalised list of SOPs that are needed. This builds on the draft list you provided earlier.

Additional Reading:

How to Save Money with SOPs that Stand the Test of Time

15 – Protocol Quality Procedures

Signature log:

Protocol Quality Procedures GetReskilled

In this section you summarize a signature-log procedure. This will be used to create the log contained in the Appendix.

How this would “look” in the real world:

When you first begin working in a Validation Team, you will simply be asked to sign this Signature Log checksheet. As you gain seniority in the team, you will be asked to explain what is included in the checksheet and the procedure for defining who is eligible to sign off documents in the protocol

Every individual who has made an entry into the protocol needs to complete the Signature Log Checksheet. This Signature Log Checksheet summarises for an auditor who the key people were in validating this piece of equipment.

It also acts as a summary of their professional experience for an auditor, as well as providing a way of tracking who works on what projects. Remember, a Validation Team on a project such as an installation and verification validation protocol could be made up of both in-house staff and outside contractors. You need to know what company they work for, what their subject matter expertise is, whether they are GMP trained, whether they are adhering to their roles and responsibilities etc.

Deviation procedure:

Protocol Quality Procedures GetReskilled

In this section you summarize a protocol deviation procedure. This will be used to create the deviation form contained in the Appendix.

How this would “look” in the real world:

There will always be deviations. Initially your task will mostly just involve filling up deviation checksheets and giving them to your boss to resolve with the relevant Subject Matter Experts (SME’s).

As you move up the career ladder, and you begin writing protocols and test procedures of your own, your job will be to find these deviations, and resolve them in real-time with the relevant Subject Matter Experts.

An FDA auditor is not looking for the deviation section of the protocol to be blank, in fact that worries them a bit (as it could mean you did not complete a thorough investigation of the system). What an FDA auditor is looking for in the deviation checksheets is that when you found a problem, you followed procedure, contacted the relevant people and solved the problem as quickly as possible.

There are categories used to describe different deviations. The 5 categories are:

  1. Correction of formatting and typographical errors (protocol generation errors)
  2. Data entry errors during execution
  3. Repeating a test due to an identifiable technical error during execution
  4. Making a test criteria less stringent
  5. Deletion of a test

Change control procedure:

Protocol Quality Procedures GetReskilled

In this section you summarize a change control procedure. This will be used to create the change control form contained in the Appendix.

How this would “look” in the real world:

When you discover more serious deviations, you and your team may need to make a change to the detail drawings or the design to resolve the problem. This will lead to you having to follow the “Change Control Procedure”.

For example, if you have identified any category 3, 4, or 5 deviations, you will most likely have to complete a change control procedure form as well. Basically – you’ve identified a problem (e.g. you fail to get the expected results from a functional test), how do you suggest it is solved?

When you start in a Validation Team, your role will stop at completing the detailed deviation form. However, as you gain seniority, you will also be expected to be part of the change control process and maybe eventually even lead the change control committee (which is made up of Subject Matter Experts from across the relevant departments). So this is important to learn now.

There are 7 steps to the change control process:

  1. Describe the proposed change
  2. Document and justify the change
  3. Evaluate the risks and impact of the change
  4. Change control committee will accept or reject the request for change
  5. Develop and verify the change
  6. Approve and implement the change
  7. Close the change and make sure all documentation is in order.

16 – P&ID Walkdown Installation Verification and Validation Protocol GMP Checksheet

P&ID-Walkdown-Installation-Verification-GetReskilled

How this would “look” in the real world:

A key inspection of any piece of equipment is a visual inspection. This will probably be one of the first tasks you complete on a Validation Team.

You would be sent out onto the factory floor (sometimes in teams of two or more) and visually confirm that what has been installed matches what’s drawn on the P&ID.

You’d then check that each part of the system has been correctly labelled (i.e. if a pipe on the P&ID has been tagged “P-001”, the same pipe on the factory floor should have a label “P-001” written on it). This walkdown only looks at what is on the factory floor and compares it to what is in the P&ID.

Remember: There is an installation verification testing procedure in Section 13 which reminds you of

  • Description of Test
  • Linkage to Requirements Challenged
  • Acceptance Criteria
  • Prerequisites and/or Assumptions
  • Test Method
  • Expected Results and Actual Results

This testing procedure will already have been written by the lead Validation Engineer so when you start your career, your main job will be to read and understand what’s required, then carry out the tests according to the method described, and record the results in the appropriate checksheets.

17 – Equipment Installation Verification and Validation Protocol GMP Checksheet

Equipment Installation Verification GetReskilled
  1. Relate the Equipment List to the P&ID and the URS
  2. Prepare installation verification check sheets for the necessary equipment items by pre-populating details for Manufacturer, Capacity, M.O.C.,
  3. Design Pressure Range and Design Temp Range.

You would then go out and check the details of each piece of equipment in person

You should only be validating the equipment that you have decided is operationally critical components and so if the items you complete checksheets for, are not the items in your summary table, an FDA auditor is going to have questions.

How this would “look” in the real world:

This is one of the first jobs you will do after you join a Validation Team.

There is an installation verification testing procedure in Section 13 which reminds you of

  • Description of Test
  • Linkage to Requirements Challenged
  • Acceptance Criteria
  • Prerequisites and/or Assumptions
  • Test Method
  • Expected Results and Actual Results

This testing procedure will already have been written by the lead Validation Engineer so when you start your career, your main job will be to read and understand what has to be carried out, then carry out the tests according to the method described and record the results in the appropriate checksheets.

If you find any deviations, make sure to follow your company’s deviation procedure and complete the relevant checksheet for deviations.

18 – Instrumentation Installation Verification and Validation Protocol GMP Checksheet

18 Instrumentation-Installation-Verification-GetReskilled
  1. Prepare installation verification check sheets for necessary instrument items. Make sure to read and confirm that the “Instrument Installation Verification” check sheets cover all the intent and procedures outlined in the “Instrument Installation Verification Testing Procedure”
  2. Prepare these instrument installation verification check sheet sequences; remember to pre-populate details, where available, for Manufacturer, Calibrated Range and MOC on Part-1 of the checksheet sequence.
  3. You would then go out and check the details of each instrument in person

You should only be validating the equipment that you have decided is operationally critical components and so if the items you complete checksheets for, are not the items in your summary table, an FDA auditor is going to have questions.

How this would “look” in the real world:

This is one of the first jobs you will do after you join a Validation Team.

There is an installation verification testing procedure in Section 13 of the protocol, which reminds you of

  • Description of Test
  • Linkage to Requirements Challenged
  • Acceptance Criteria
  • Prerequisites and/or Assumptions
  • Test Method
  • Expected Results and Actual Results

This testing procedure will already have been written by the lead Validation Engineer so when you start your career, your main job will be to read and understand what has to be carried out, then carry out the tests according to the method described and record the results in the appropriate checksheets.

If you find any deviations, make sure to follow your company’s deviation procedure and complete the relevant checksheet for deviations.

19 – Piping Installation Verification and Validation Protocol GMP Checksheet

Piping-Installation-Verification-GetReskilled

Complete the piping installation verification checksheets.

  1. Make sure to read and confirm that the “Piping Installation Verification” check sheets cover all the intent and procedures outlined in the “Piping Installation Verification Testing Procedure”.
  2. Prepare installation verification check sheets for the individual pipes and note the inline component item numbers and descriptions associated with each line in the ‘Comments’ box at the bottom of the sheet. Pre-populate the relevant piping installation verification check sheet sequence with the Tag # and material-of-construction (MOC) details.
  3. Locate each piece of piping on-site and complete the Piping Installation Verification check sheets for the instrument items in the field.

How this would “look” in the real world:

This is one of the first jobs you will do after you join a Validation Team.

There is an installation verification testing procedure in Section 11.4 of the protocol, which reminds you of

Description of Test
Linkage to Requirements Challenged
Acceptance Criteria
Prerequisites and/or Assumptions
Test Method
Expected Results and Actual Results

This testing procedure will already have been written by the lead Validation Engineer so when you start your career, your main job will be to read and understand what has to be carried out, then carry out the tests according to the method described and record the results in the appropriate checksheets.

If you find any deviations, make sure to follow your company’s deviation procedure and complete the relevant checksheet for deviations.

20 – Traceability Matrix

Traceability Matrix GetReskilled

A Traceability Matrix is a tool used to ensure that all requirements defined within the User Requirement Specifications (URS) are tested in the test protocols. The traceability matrix is a tool both for the validation team (to ensure that requirements are not lost during the validation project) and for auditors (to review the validation documentation).

How this would “look” in the real world:

The purpose of a traceability matrix is to help plan and track the testing programme. When you begin in a Validation Team, you are not going to be expected to complete this. However, this will become more and more important as you gain seniority in the team.

This would be a “live” part of the protocol when test strategies are just being developed – it is a very straightforward way of tracking the different tests you want to perform on the system.

6 Things to Keep in Mind when Writting a Validation Protocol

  1. In a real world setting, especially at the beginning of a pharmaceutical validation career, you’ll be working within a larger team to complete validation activities.
  2. Different companies use different templates, different titles, and have their own processes for conducting validation activities. The purpose of this example isn’t to show you an exhaustive account of running a validation procedure, it’s to show you how completed documentation is built up in a systematic way, making sure you understand the processes at each stage.
  3. As a Validation Team Member, your most important skills are your soft skills. Technical skills can be taught – soft skills are instinctive and are what will set you apart at the end of the day. A Validation Team Member has good written communication skills, self motivation, problem solving skills, organisation skills, IT skills, and great attention to detail.
  4. You’ll notice when you start the protocol, that there are two parts to it. There is the written section at the front, and the checksheets in the back (the appendices). The front section where there are summary descriptions that describe what is written on the checksheets. An auditor will always read the front of the protocol first and only turn to the appendices when they have a question or see a problem with the information provided. This means it is really important that what is in both parts match and the summaries you write in the front are representative of the results on the checksheets in the appendices.
  5. While you are completing this protocol, you will notice that all the sections are interlinked. No one section sits in isolation from the others. Each section builds on what has been written before or what will be written later in the protocol.
  6. You will also notice that you don’t complete it in page order. Instead you will be moving around the protocol a bit. Most of the protocol is completed in the order in which it is laid out, however, every now and again you will need to go to an appendix to get information or complete a task before going back to an earlier section. This is the way protocols will be laid out in the field as they are written for a person or an auditor who is coming with no knowledge of the system to be able to understand what was carried out to deem a system fit-for-intended-use.

You might also be interested in reading:

About the Author

Photo of GetReskilled President Gerry Creaner

Gerry Creaner

President
Senior Lecturer with GetReskilled

Gerry Creaner has over 30-years of experience in the Life Sciences Manufacturing industry across a range of technical, managerial and business roles. He established a very successful engineering consultancy prior to founding GetReskilled, an online education and learning business, with offices in Singapore, Ireland and Boston (USA), focussed on the manufacture of safe and effective medicines for the public.

He is also a founding Director of two Singapore based philanthropic organizations, the Farmleigh Fellowship and the Singapore-Ireland Fund, both of which deepen the well established and historical Singapore – Ireland relationship and deliver long-term benefits to both countries.

Gerry has an undergraduate degree in Chemical Engineering (UCD, 1980) and an MSc (Management) from Trinity College Dublin (2003) and is currently doing research for his Ph.D.

Our Team

Donagh Fitzgerald

Head of Marketing & Product Development
Mechanical/Production Engineer

Donagh looks after the marketing and product development including the training and pedagogical elements of our programs and makes sure that all GetReskilled’s users can have a great online learning experience. Donagh has lived and worked in many countries including Ireland, America, the UK, Singapore, Hong Kong and Japan. Donagh has also served as the Program Manager for the Farmleigh Fellowship based out of Singapore.

Donagh holds Degrees in Production Engineering and Mechanical Engineering from South East Technological University, Ireland.